• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ALTRUA 40; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ALTRUA 40; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number S404
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
Health Effect - Clinical Code Tachycardia (2095)
Date of Event 11/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned device was inspected and analyzed upon receipt at our post market quality assurance laboratory.The issue was caused by interactions between the device and hospital monitoring or diagnostic equipment when the minute ventilation feature is programmed on.Pacing and sensing functions were tested and the device was verified to operate as expected.
 
Event or Problem Description
It was reported that the patient with this pacemaker was noted to have a high pacing rate.The patient had previously been put on a life vest after a having experienced a heart attack.The field representative programmed the minute ventilation (mv) sensor off.Boston scientific technical services (ts) discussed that electromagnetic interference from the vest was likely affecting the mv signal and by programming the mv sensor off, the pacing from the device was affected.Ts discussed how equipment such as the vest can affect mv signal.The device remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRUA 40
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14067543
Report Number2124215-2022-03942
Device Sequence Number4766055
Product Code LWP
UDI-Device Identifier00802526484452
UDI-Public00802526484452
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
N970003/S106
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/07/2011
Device Model NumberS404
Device Catalogue NumberS404
Device Lot Number592512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2022
Initial Report FDA Received Date04/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age83 YR
Patient SexFemale
-
-