• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PROPONENT EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PROPONENT EL DR; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number L221
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this pacemaker was depleting faster than expected.When implanted in 2017, the device had an estimated longevity based on programming parameters of 11.5 years but there is currently 11 months of estimated longevity remaining only 5 years later.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
The product has been received for analysis.This report will be updated upon completion of analysis.
 
Event or Problem Description
It was reported, that this pacemaker was depleting faster than expected.When implanted in 2017, the device had an estimated longevity, based on programming parameters of 11.5 years, but there is currently 11 months of estimated longevity remaining only 5 years later.No adverse patient effects were reported.New information received, indicates that this device was explanted and returned for analysis.
 
Event or Problem Description
It was reported that this pacemaker was depleting faster than expected.When implanted in 2017, the device had an estimated longevity based on programming parameters of 11.5 years but there is currently 11 months of estimated longevity remaining only 5 years later.No adverse patient effects were reported.New information received indicates that this device was explanted and returned for analysis.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPONENT EL DR
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14067951
Report Number2124215-2022-07013
Device Sequence Number12005381
Product Code LWP
UDI-Device Identifier00802526559112
UDI-Public00802526559112
Combination Product (Y/N)N
Initial Reporter CountryNU
PMA/510(K) Number
N970003/S167
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/19/2017
Device Model NumberL221
Device Catalogue NumberL221
Device Lot Number706810
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/20/2022
Supplement Date Received by Manufacturer05/06/2022
05/31/2022
Initial Report FDA Received Date04/09/2022
Supplement Report FDA Received Date05/22/2022
06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/12/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexFemale
-
-