| Model Number |
L221 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable device was suspected to be exhibiting premature battery depletion behavior.A decrease in the estimated battery longevity was observed during a recent follow-up visit.Subsequently, this device was explanted and replaced.No additional adverse patient effects were reported.This device is expected for return.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable device was suspected to be exhibiting premature battery depletion behavior.A decrease in the estimated battery longevity was observed during a recent follow-up visit.Subsequently, this device was explanted and replaced.No additional adverse patient effects were reported.This device was received for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of compromised capacitors.This resulted in the observed premature battery depletion.It was determined that the capacitors were compromised due to the presence of excess hydrogen in the device case.Boston scientific issued a field safety notice regarding a subset of devices in the accolade pacemaker family that have a potential of exhibiting this behavior.This particular device is included in the hydrogen induced premature depletion advisory population.
|
| |
|
Search Alerts/Recalls
|