BOSTON SCIENTIFIC CORPORATION TELIGEN; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number E110
Device Problem Delayed Charge Time (2586)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2019
Event Type  Injury  
Manufacturer Narrative
If additional information is received, a supplemental report will be filed at that time.
 
Event Description
It was reported that the patient with this implantable cardioverter defibrillator (icd) presented to the emergency room after receiving shocks.The device displayed a code 1007 which is indicative of capacitor charge time has exceeded the limit.Subsequently, the device was explanted and replaced.The new device remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELIGEN
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14069014
MDR Text Key288995891
Report Number2124215-2022-09216
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960040/S155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/17/2012
Device Model NumberE110
Device Catalogue NumberE110
Device Lot Number167862
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/20/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age51 YR
Patient SexFemale
-
-