Catalog Number RBYPODJ45
Medical Device Problem Codes Physical Resistance/Sticking (4012); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the celiac trunk and splenic artery using a pod packing coil (podj), a lantern delivery microcatheter (lantern) and a non-penumbra guide catheter.During the procedure, after advancing approximately twenty-five percent of the podj into the aneurysm using the lantern and guide catheter, the physician decided to retract the podj.While retracting the podj, the physician experienced resistance and subsequently, the podj unintentionally detached.Therefore, the physician removed the guide catheter with the lantern that contained the detached podj and then flushed the podj out of the lantern.The procedure was completed using a new podj and the same lantern.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the following section is being updated based on additional information provided by the penumbra distributor on 13-apr-2022: 1.Section b.Box 5.Describe event or problem h3 other text : placeholder.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the celiac trunk and splenic artery using a pod packing coil (podj), a lantern delivery microcatheter (lantern) and a non-penumbra guide catheter.During the procedure, after advancing approximately twenty-five percent of the podj into the aneurysm using the lantern and guide catheter, the podj would not advance further.Therefore, the physician decided to retract the podj.While retracting the podj, the physician experienced resistance and subsequently, the podj unintentionally detached.Therefore, the physician removed the guide catheter with the lantern that contained the detached podj and then flushed the podj out of the lantern.The procedure was completed using a new podj and the same lantern.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD PACKING COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key14069062
Report Number3005168196-2022-00166
Device Sequence Number4641925
Product Code HCG
UDI-Device Identifier00814548016726
UDI-Public00814548016726
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K141134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/27/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberRBYPODJ45
Device Lot NumberF102894
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2022
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date04/09/2022
Supplement Report FDA Received Date04/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-