Model Number D001
Medical Device Problem Codes Premature Discharge of Battery (1057); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that error code 1007 and 1003 were displayed since the last device follow up.A review of the device data by technical services (ts) noted that the since the device had triggered end of life (eol) there was little diagnostic data.Ts confirmed the error codes and noted that fault code 1007 indicated that the capacitor charge time was exceeded.The last capacitor reform was (b)(6) 2021 and the charge time exceeded 45 seconds, after which the device aborts further charging.Ts also noted that since the device triggering fc1003 the battery was depleting more quickly than expected.Immediate device replacement was recommended.The device remains implanted.The patient remained hospitalized and no further adverse patient effects were reported.
 
Additional Manufacturer Narrative
The device remains implanted.Should additional information become available this investigation will be updated.
 
Event or Problem Description
It was reported that error code 1007 and 1003 were displayed since the last device follow up.A review of the device data by technical services (ts) noted that the since the device had triggered end of life (eol) there was little diagnostic data.Ts confirmed the error codes and noted that fault code 1007 indicated that the capacitor charge time was exceeded.The last capacitor reform was march 15, 2021 and the charge time exceeded 45 seconds, after which the device aborts further charging.Ts also noted that since the device triggering fc1003 the battery was depleting more quickly than expected.Immediate device replacement was recommended.Additional information noted that at the last telemetry follow up in march 2022 it was not possible to estimate the battery due to alleged battery depletion.The device was explanted, replaced, and is pending a return.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
The device was planned to be returned.Should the device be returned analysis will be performed and this investigation will be updated.
 
Event or Problem Description
It was reported that error code 1007 and 1003 were displayed since the last device follow up.A review of the device data by technical services (ts) noted that the since the device had triggered end of life (eol) there was little diagnostic data.Ts confirmed the error codes and noted that fault code 1007 indicated that the capacitor charge time was exceeded.The last capacitor reform was (b)(6) 2021 and the charge time exceeded 45 seconds, after which the device aborts further charging.Ts also noted that since the device triggering fc1003 the battery was depleting more quickly than expected.Immediate device replacement was recommended.Additional information noted that at the last telemetry follow up in march 2022 it was not possible to estimate the battery due to alleged battery depletion.The device was explanted, replaced, and was received for analysis.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.External visual inspection of the device revealed no anomalies.Review of the device memory confirmed that a low voltage alert was recorded.The device case was then opened to facilitate analysis of the internal components.The battery was separated from the other device electronics and the overall current draw of the circuitry was measured.A normal current drain was observed within the circuitry.Collectively, the pattern of irregular daily battery voltage measurements in conjunction with normal power levels and device hybrid current draw is consistent with behavior of devices where a latent current leakage path has occurred within the battery itself, resulting in a partial depletion of the battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORIGEN MINI ICD VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14069122
Report Number2124215-2022-08336
Device Sequence Number4766092
Product Code LWS
UDI-Device Identifier00802526533723
UDI-Public00802526533723
Combination Product (Y/N)N
Initial Reporter CountryBK
PMA/510(K) Number
P960040/S306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/30/2017
Device Model NumberD001
Device Catalogue NumberD001
Device Lot Number102118
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/18/2022
Supplement Date Received by Manufacturer04/13/2022
04/27/2023
Initial Report FDA Received Date04/10/2022
Supplement Report FDA Received Date05/25/2022
05/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-