BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24700
Medical Device Problem Codes Material Puncture/Hole (1504); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that balloon rupture and shaft perforation occurred.The 100% stenosed target lesion was located in the moderately tortuous and moderately calcified vessel below the knee.A 4.0mmx100mmx80cm (4f) sterling balloon catheter was advanced for dilatation.However, during multiple times of inflation at approximately 10 atmospheres, the balloon ruptured.Additionally, a wire came out through the shaft at the balloon part.The procedure was completed with a different device.There were no patient complications nor injuries reported.
 
Event or Problem Description
It was reported that balloon rupture and shaft perforation occurred.The 100% stenosed target lesion was located in the moderately tortuous and moderately calcified vessel below the knee.A 4.0mmx100mmx80cm (4f) sterling balloon catheter was advanced for dilatation.However, during multiple times of inflation at approximately 10 atmospheres, the balloon ruptured.Additionally, a wire came out through the shaft at the balloon part.The procedure was completed with a different device.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
(e1) initial reporter address 1: (b)(6).Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.Microscopic inspection showed a kink of the inner tube 3 cm from the tip.The hole was next to the kink and looked consistent with damage caused by a guidewire perforating the inner tube.A rupture of the balloon was not confirmed but the device did have liquid come out the distal end due to the hole of the inner tube.Device analysis determined the condition of the returned device was consistent with the reported information of a hole/perforation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14069262
Report Number2134265-2022-03351
Device Sequence Number4766117
Product Code DQY
UDI-Device Identifier08714729123514
UDI-Public08714729123514
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K053116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/21/2024
Device Model Number24700
Device Catalogue Number24700
Device Lot Number0027184838
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer05/06/2022
Initial Report FDA Received Date04/10/2022
Supplement Report FDA Received Date05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-