| Model Number |
24700 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture and shaft perforation occurred.The 100% stenosed target lesion was located in the moderately tortuous and moderately calcified vessel below the knee.A 4.0mmx100mmx80cm (4f) sterling balloon catheter was advanced for dilatation.However, during multiple times of inflation at approximately 10 atmospheres, the balloon ruptured.Additionally, a wire came out through the shaft at the balloon part.The procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture and shaft perforation occurred.The 100% stenosed target lesion was located in the moderately tortuous and moderately calcified vessel below the knee.A 4.0mmx100mmx80cm (4f) sterling balloon catheter was advanced for dilatation.However, during multiple times of inflation at approximately 10 atmospheres, the balloon ruptured.Additionally, a wire came out through the shaft at the balloon part.The procedure was completed with a different device.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(e1) initial reporter address 1: (b)(6).Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.Microscopic inspection showed a kink of the inner tube 3 cm from the tip.The hole was next to the kink and looked consistent with damage caused by a guidewire perforating the inner tube.A rupture of the balloon was not confirmed but the device did have liquid come out the distal end due to the hole of the inner tube.Device analysis determined the condition of the returned device was consistent with the reported information of a hole/perforation.
|
| |
|