| Model Number |
SSFT215PRE |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not available to manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure the tip of the guidewire (subject device ) broke.There had been some wire reshaping due to difficult anatomy.There was difficulty advancing the guidewire (subject device) but it went into the aneurysm for the most part.Also, there was some resistance when pulling the guidewire (subject device) out of the aneurysm.Possible that the guidewire (subject device) hooked a little in the stent.Then the distal part of the guidewire (subject device) was left in the microcatheter and this had to be pulled out to the intermediate catheter with the guidewire (subject device) and a medical intervention was performed as this had to secure it with a stent retriever when it seemed to drop.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
There are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer indicated that no anomalies were noted to the device during initial inspection and preparation, the device was prepared as per the dfu, a non-stryker microcatheter was used with the subject guidewire, the guidewire was reshaped due to difficult anatomy, continuous flush was maintained for the duration of the procedure, and friction/resistance was noted when the wire was attempted to be removed from the aneurysm.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned to this complaint.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure the tip of the guidewire (subject device ) broke.There had been some wire reshaping due to difficult anatomy.There was difficulty advancing the guidewire (subject device) but it went into the aneurysm for the most part.Also, there was some resistance when pulling the guidewire (subject device) out of the aneurysm.Possible that the guidewire (subject device) hooked a little in the stent.Then the distal part of the guidewire (subject device) was left in the microcatheter and this had to be pulled out to the intermediate catheter with the guidewire (subject device) and a medical intervention was performed as this had to secure it with a stent retriever when it seemed to drop.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
|
| |
|