Model Number SSFT215PRE
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is not available to manufacturer.
 
Event or Problem Description
It was reported that during the procedure the tip of the guidewire (subject device ) broke.There had been some wire reshaping due to difficult anatomy.There was difficulty advancing the guidewire (subject device) but it went into the aneurysm for the most part.Also, there was some resistance when pulling the guidewire (subject device) out of the aneurysm.Possible that the guidewire (subject device) hooked a little in the stent.Then the distal part of the guidewire (subject device) was left in the microcatheter and this had to be pulled out to the intermediate catheter with the guidewire (subject device) and a medical intervention was performed as this had to secure it with a stent retriever when it seemed to drop.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
There are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer indicated that no anomalies were noted to the device during initial inspection and preparation, the device was prepared as per the dfu, a non-stryker microcatheter was used with the subject guidewire, the guidewire was reshaped due to difficult anatomy, continuous flush was maintained for the duration of the procedure, and friction/resistance was noted when the wire was attempted to be removed from the aneurysm.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned to this complaint.
 
Event or Problem Description
It was reported that during the procedure the tip of the guidewire (subject device ) broke.There had been some wire reshaping due to difficult anatomy.There was difficulty advancing the guidewire (subject device) but it went into the aneurysm for the most part.Also, there was some resistance when pulling the guidewire (subject device) out of the aneurysm.Possible that the guidewire (subject device) hooked a little in the stent.Then the distal part of the guidewire (subject device) was left in the microcatheter and this had to be pulled out to the intermediate catheter with the guidewire (subject device) and a medical intervention was performed as this had to secure it with a stent retriever when it seemed to drop.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHRO SELECT-14 SOFT PS 215CM
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer (Section G)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key14069928
Report Number3012931345-2022-00078
Device Sequence Number2571702
Product Code MOF
UDI-Device Identifier07613327508598
UDI-Public07613327508598
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
K202522
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/08/2024
Device Model NumberSSFT215PRE
Device Catalogue NumberSSFT215PRE
Device Lot Number0000077797
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/05/2022
Supplement Date Received by Manufacturer07/06/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-