• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE,16F; FEEDING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE,16F; FEEDING DEVICE Back to Search Results
Model Number 000716
Medical Device Problem Codes Break (1069); Fluid/Blood Leak (1250); Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that next day post feeding tube placement procedure, the catheter allegedly came out and had leak issue.It was further reported that the cuff allegedly found to be ruptured.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has not been returned to the manufacturer for evaluation.However, a photo and a video was provided for review.The investigation of the reported event is currently underway.(expiry date: 04/2023).
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical sample was not returned for evaluation, one electronic photo and one electronic video was returned for evaluation.The video shows the clinician unpacking one tri-funnel g tube from the package.The device was able to be inflated without any issue.The photo shows the distal portion of one tri-funnel g tube.A break was noted on the inflation balloon of the device.The investigation is confirmed for the reported break and fluid leak issues as a break was noted on the balloon of the returned device.However, the investigation is inconclusive for the reported migration issue as the exact circumstances at the time of the reported event cannot be verified.A definitive root cause for the reported failure could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 04/2023).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that next day post feeding tube placement procedure, the catheter allegedly came out and had leak issue.It was further reported that the cuff allegedly found to be ruptured.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE,16F
Common Device Name
FEEDING DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14071242
Report Number3006260740-2022-01220
Device Sequence Number6020029
Product Code KNT
UDI-Device Identifier00801741037139
UDI-Public(01)00801741037139
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K063118
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number000716
Device Catalogue Number000716
Device Lot NumberNGDX4567
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer05/25/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date05/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-