| Model Number |
502-11-54E |
| Medical Device Problem Codes |
Loss of Osseointegration (2408); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
03/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
|
| |
|
Event or Problem Description
|
|
Trident cup was loose in bone, and turned to an undesirable position.
|
| |
|
Event or Problem Description
|
|
Trident cup was loose in bone, and turned to an undesirable position.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding loosening involving a trident shell was reported.The event was not confirmed.Method & results: -product evaluation and results: visual inspection: visual inspection of the returned device indicated that the device has residue on the outer shell consistent with time spent implanted and minor scratches consistent with explantation.There is nothing further remarkable to note.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to loosening of the cup.Visual inspection of the returned device indicated that the device has residue on the outer shell consistent with time spent implanted and minor scratches consistent with explantation.There is nothing further remarkable to note.The exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Search Alerts/Recalls
|