SMITH & NEPHEW, INC. EVOS SMALL 2.5MM DRILL W/AO QC LONG; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 71175022
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the evos small 2.5mm drill w/ao qc long was broken and the surgeon decided to leave it inside.The procedure was completed, with a non-significant delay, using a s+n back-up device.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during an open reduction in an internal fixation of tibial platu, an evos small 2.5mm drill w/ao qc long broke and the surgeon decided to leave it inside the patient.The procedure was completed, with a non-significant delay, using a s+n back-up device.Patient was not harmed beyond the problem reported.
 
Additional Manufacturer Narrative
B4, h3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, this case reports a drill bit breakage of the evos 2.5mm long drill during an open reduction internal fixation tibia plateau.According to the report, the surgeon made the decision to leave the broken drill bit inside of the patient.No clinically relevant supporting documentation was provided for inclusion in this medical investigation.Consequently, the clinical root cause of the reported drill bit breakage could not be determined.Per subsequent e-mail, no additional surgery is required.The evos 2.5mm long drills are comprised of 17.4 ph ss, they are manufactured and intended as externally communicating devices and are not approved for long term internal tissue exposure and long-term implantation data is not available.Although we cannot make conclusions on the patient impact of the non-implantable foreign body, we cannot rule out the possibility of micro-motion and/or migration, local skin irritation and discomfort.Based on the information provided, the procedure was completed with a non-significant delay, using a s+n back-up device.According to the report, the patient was not harmed as consequence of this problem.Therefore, no further medical assessment can be rendered at this time.Should any additional relevant clinical information be provided, this case would be re-assessed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL 2.5MM DRILL W/AO QC LONG
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14072042
Report Number1020279-2022-01749
Device Sequence Number11424844
Product Code HWE
UDI-Device Identifier00885556635766
UDI-Public00885556635766
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71175022
Device Catalogue Number71175022
Device Lot Number21JUQ0026
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer05/13/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date06/07/2022
Date Device Manufactured10/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-