BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL TRACKER; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10014
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).The sample is under evaluation by manufacturing site.
 
Event or Problem Description
It was reported that, during cori assisted tka surgery, the robotic drill kept coming in and out of view with the camera.They wiped the flat markers off and ensured they were seated, but still experienced the drill coming in and out of view with the camera even though it was in a direct line of site with the camera.Upon visual inspection, it appeared that the real intelligence robotic drill tracker was bent.Procedure was resumed, after a non-significant delay, using the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Section h10: the real intelligence robotic drill tracker, rob10014, lot 20fumc001 used for treatment was returned for evaluation.A relationship between the reported event and the device was established.The reported problem was visually confirmed.There is slight appearance of one bent flat marker connector.Most prominent when flat marker was placed on the connector.A functional evaluation was performed.The reported problem was confirmed.A case was created.Drill tracker array would float in and out of view.The most likely cause of this event is the connector becomes bent as a result of tight fitting flat markers and the need for a tool to remove.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of the reported complaint.Care and caution should be exercised during surgical assembly and use of the drill tracker array.Refer to the cori surgical system user's manual for proper handling.This failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTIC DRILL TRACKER
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14072119
Report Number3010266064-2022-00278
Device Sequence Number2632198
Product Code OLO
UDI-Device Identifier00885556757338
UDI-Public00885556757338
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberROB10014
Device Catalogue NumberROB10014
Device Lot Number20FUMC001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer04/25/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date04/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CORI ROBOTICS USA-(B)(4)
Patient SexUnknown
-
-