Product event summary: the introducer was returned and analyzed.The inner/sterile package (extrinsic) had damaged trays.The delivery system introducer sheath was kinked/buckled.The delivery system introducer dilator was kinked/buckled.The analyst noted the introducer was returned in its unopened sealed inner packaging with a use by date of 2024-02-09.The outer packaging was not returned.The inner tray inserts were not fastened in their locations on the inner tray to hold in the introducer sheath and dilator.The inner tray was damaged.The introducer sheath was kinked at 45 cm.The shaft of the introducer dilator was bent at 31.5 cm.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|