MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND SCREW NO LOCK D4.5X30MM; DELTA XTEND IMPLANTS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND SCREW NO LOCK D4.5X30MM; DELTA XTEND IMPLANTS : SCREWS Back to Search Results
Model Number 130770030
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 03/18/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that surgeon had implanted the metaglene with screws.He mentioned that the patient bone was very soft and screws were not getting great purchase.While implanting the glenosphere, he was having trouble getting it to engage in the metaglene.When he removed the glenosphere and noticed the metaglene was toggling.He removed the metaglene and the glenoid was fractured.He decided to convert to a hemi arthroplasty.Since they did not have instruments in the hospital, he used another company that had instruments sterile on the shelf.Doe: (b)(6) 2022, affected side: left shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW NO LOCK D4.5X30MM
Common Device Name
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14073253
Report Number1818910-2022-06457
Device Sequence Number2570215
Product Code PHX
UDI-Device Identifier10603295027805
UDI-Public10603295027805
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K120174
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number130770030
Device Catalogue Number130770030
Device Lot Number5378234
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2022
Supplement Date Received by Manufacturer04/20/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date04/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DXTEND METAGLENE; DXTEND SCREW LOCK D4.5X24MM; DXTEND SCREW NO LOCK D4.5X18MM; DXTEND SCREW NO LOCK D4.5X18MM; DXTEND SCREW NO LOCK D4.5X18MM; DXTEND SCREW NO LOCK D4.5X30MM; XTNDGLENO ECCD38MM+0MM SHRTPST
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-