• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CEMENTED STEM 16X80MM; KNEE STEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CEMENTED STEM 16X80MM; KNEE STEMS Back to Search Results
Model Number 1512-16-080
Medical Device Problem Codes Loss of or Failure to Bond (1068); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/26/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Clinical adverse event received for aspiration of study knee (right) - confirmation of loosening tibial component.Event is serious and is considered severe.Event is possibly related to device and is probably related to procedure.Date of implantation: (b)(6) 2018, date of event (onset): (b)(6) 2021, (right knee).Treatment: aspiration of knee.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Clinical notification update ((b)(4) clinical update ad 29 mar 2022) reviewed, along with the current clinical database.The diagnosis as written in the clinical database adverse event log entry is "aspiration of study knee - as a result of raised crp and infection - confirmation of loosening tibial component".This indicates that the infection has been confirmed.There has yet to be any reported knee revision, so the tibial tray loosening remains unconfirmed.Therefore, impacted products, and patient coding will be updated to address this new information.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional information was received indicating that the "bone cement has leaked from beneath the tray onto the sleeve during insertion / pressurization which prevented biological ingrowth.".
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this complaint was not received.Removal off the poly plug from the returned device revealed evidence that a stem was used, however is was not returned for investigation.Depuy synthese considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CEMENTED STEM 16X80MM
Common Device Name
KNEE STEMS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14073467
Report Number1818910-2022-06469
Device Sequence Number14995535
Product Code JWH
UDI-Device Identifier10603295043010
UDI-Public10603295043010
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1512-16-080
Device Catalogue Number151216080
Device Lot NumberHB2195
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2022
Supplement Date Received by Manufacturer05/12/2022
07/26/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date05/19/2022
07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/20/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CEMENTED STEM 16X80MM; ATTUNE CRS FEMORAL RT SZ 4 CEM; ATTUNE MEDIAL DOME PAT 38MM; ATTUNE POS FEM AUG SZ 4 4MM; ATTUNE PS RP INSRT SZ 4 12MM; ATUN TIB SLV M/L 29MM HALF POR; ATUNE REV RP TIB BASE SZ 4 CEM; PALACOS R+G 2X40 CAT.NO.66017569 LOT 88664782; PALACOS R+G 2X40 CAT.NO.66017569 LOT 88664782
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
Patient Weight89 KG
-
-