| Model Number |
N/A |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
03/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that patient underwent a hip revision approximately 1.5 years post implantation due to thigh pain and instability.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Implanted in (b)(6) 2020.Concomitant medical products: 51-130110 ¿ taperloc stem ¿ 6616845.Unknown cup ¿ unknown part and lot.Unknown biolox head ¿ lot 3108029.Report source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for the investigation as the hospital did not approve of the return.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.A review of the available records identified findings of the reported issues thigh pain and instability as reported.A review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|