Product complaint #: (b)(4).Component code: (b)(4) device not returned.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The following information was requested, but unavailable: did the event occur during one or multiple patient procedures? what is the total number of procedures? have any of these events been previously reported to ethicon? if so, provide the respective reference number(s).If this event occurred in multiple procedures, please create a product complaint for each event and provide the respective reference number(s).The single complaint was reported with multiple events.There are no additional details regarding the additional events.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.Event related to mw # 2210968-2022-02553.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint # (b)(4).Date sent to the fda: 5/3/2022 additional information: d9, h3, h6 h6 component code: g07002 no device problem found h3 investigational summary: one empty overwrap, two empty labeled winding former, and two needle-suture pieces product code 8695h were returned to ethicon for analysis.Visual inspection was conducted on the returned device.In order to evaluate the conditions of the returned sample, the swage area and hole of the needle were noted with marks that appear to be by a surgical instrument: also, the suture pieces were observed with body fluids and the end was cut.The condition of the sample received indicates improper handling of the device.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: one opened box with thirty-one packets of product code 8695h were returned to ethicon for analysis with the packaging closed.Upon initial inspection to the sample no external damage was observed on the packet.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected and no damages, breakage on the body, tip, or swage area were observed.The suture was dispensed without problems and examined along of the strand to detect any issue related to damaged or breakage suture and no defects were observed during evaluation.Functional tests were performed, to needle by resistance and the tensile strength and met the requirements.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: one opened box with twenty-five packets of product code 8695h was returned to ethicon for analysis with the packaging closed.Upon initial inspection, of the samples, no external damage was observed on the packet.In order to evaluate the conditions of the returned samples, thirteen packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected and no damages, breakage on the body, tip, or swage area were observed.The suture was dispensed without problems and examined along the strand to detect any issue related to damaged or breakage suture and no defects were observed during evaluation.Functional tests were performed, to needle by resistance and the tensile strength and met the requirements.The event described could not be confirmed as the device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|