MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CLEO SUBQ DISPOSABLES; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CLEO SUBQ DISPOSABLES; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 21-7232-24
Medical Device Problem Codes Loss of or Failure to Bond (1068); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after 2 attempts to insert a soft set that the adhesive for the cannula/hub was not sticking to the patient.Soft set insertion piece inspected on a third unit and noted that the needle was bent.Other soft sets from the same box checked by nurse and several noted to have bent needles.We have also experienced the issue with cannula/hub not releasing from the insertion piece/not sticking to the skin/tegaderm with other units (where there was no issue with bent needles).No patient injury was reported.
 
Additional Manufacturer Narrative
A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.No product or photographic evidence were provided to aid in this investigation.The complaint report offered insufficient details to determine whether this product functioned as intended, or was used in a manner consistent with its instructions for use (ifu) or failed to meet product specifications.Lacking any additional evidence, this complaint has been closed as unconfirmed.If the product is returned, smiths medical will reopen this complaint for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEO SUBQ DISPOSABLES
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14073782
Report Number3012307300-2022-06149
Device Sequence Number17685209
Product Code FPA
UDI-Device Identifier10610586028410
UDI-Public10610586028410
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K042172
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number21-7232-24
Device Lot Number4213341
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2022
Initial Report FDA Received Date04/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-