• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US HP MBT CEM KEEL PUNCH SZ 4-7; KNEE INSTRUMENT : TIBIAL KEEL PUNCH/BULLET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US HP MBT CEM KEEL PUNCH SZ 4-7; KNEE INSTRUMENT : TIBIAL KEEL PUNCH/BULLET Back to Search Results
Model Number 9505-02-012
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the connection tabs are broken off ¿ a repeating problem reported previously on the item.No loose parts.No parts remaining in patient.There was no surgical delay.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the device associated with this report was returned for analysis.Visual examination of the device found 1 assembly tab broken.Confirming, the reported allegation.Depuy synthes considers, the investigation closed at this time.Should additional information be received, the information will be reviewed.And the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HP MBT CEM KEEL PUNCH SZ 4-7
Common Device Name
KNEE INSTRUMENT : TIBIAL KEEL PUNCH/BULLET
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14074115
Report Number1818910-2022-06492
Device Sequence Number12301799
Product Code LRY
UDI-Device Identifier10603295227328
UDI-Public10603295227328
Combination Product (Y/N)N
Initial Reporter CountryNO
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9505-02-012
Device Catalogue Number950502012
Device Lot NumberSO2026139
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2022
Supplement Date Received by Manufacturer09/13/2022
09/29/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date09/14/2022
10/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-