| Model Number |
EPH02 |
| Medical Device Problem Code |
Difficult or Delayed Activation (2577)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during an ob-gyn procedure, the device started suctioning although the activation button was not pressed.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 5/5/2022.D4 batch #: v9679l.Investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the eph02 device was received with no apparent damage.The instrument was tested for continuity and was found to be conforming.The device did activate when the buttons were depressed, but no continuous activation occurred at any time during functional testing.There were no anomalies noted with the functionality of the device.The device was disassembled to inspect the internal components, no evidence was found that could have caused the reported continuous activation.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|