Model Number M00568211
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an endovive standard peg kit push method was used during a percutaneous endoscopic gastrostomy (peg) placement procedure performed on (b)(6) 2022.During the procedure, the peg tube broke at the transition zone as it was being inserted through the patient's mouth up to the stoma site.The peg tube had not gone through the stoma and was stuck in the gastric side.It was reported that the detached portion was endoscopically retrieved.The procedure was completed with a new endovive standard peg kit push method.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that an endovive standard peg kit push method was used during a percutaneous endoscopic gastrostomy (peg) placement procedure performed on (b)(6) 2022.During the procedure, the peg tube broke at the transition zone as it was being inserted through the patient's mouth up to the stoma site.The peg tube had not gone through the stoma and was stuck in the gastric side.It was reported that the detached portion was endoscopically retrieved.The procedure was completed with a new endovive standard peg kit push method =.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6 (device codes): device code a0501 captures the reportable event of peg tube detached.Block h10: product analysis found the peg tube separated from the dilator at the barbed connector joint, and the peg tubing was not returned.The reported complaint was confirmed.Based on the condition of the returned device, engineers determined that the problem observed is likely that operational factors, such as user technique/handling, patient anatomy, and the size of the incision site that the tube was being pulled through, contributed to the reported complaint.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.It is most likely that the adverse event occurred during the procedure and the device had no influence on the event which led to the reported event.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE STANDARD PEG KIT
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14074220
Report Number3005099803-2022-01920
Device Sequence Number11424863
Product Code KNT
UDI-Device Identifier08714729285205
UDI-Public08714729285205
Combination Product (Y/N)Y
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K031538
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/16/2022
Device Model NumberM00568211
Device Catalogue Number6821
Device Lot Number0027074209
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer04/19/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-