• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTURE ANCHOR, MICRO CORKSCREW FT; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTURE ANCHOR, MICRO CORKSCREW FT; SCREW, FIXATION, BONE Back to Search Results
Model Number SUTURE ANCHOR, MICRO CORKSCREW FT
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
On 3/21/2022, it was reported by a sales representative via email that (2) ar-1318ft corkscrew suture anchors broke in half inside the patient's joint.All broken fragments were retrieved and two more ar-1318ft corkscrew suture anchors were opened; however, both of the anchors fell off the inserter and could not be reloaded back on the inserter shaft.Surgeon opened a fifth ar-1318ft corkscrew suture anchor to complete the case.This was discovered during a flexor tendon repair on (b)(6) 2022.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
As the anchors were not returned, the complaint cannot be confirmed.However, additional failure modes were identified.The anchors/suture were not returned for investigation.Upon visual evaluation it was noted that the tip on one driver had broken.The broken fragment was not returned for investigation.On the second driver it was noted that the tip was twisted.The diameter of each shaft prior to the breakage site was measured and found to be within the specifications listed in drawing c0764.The cause remains undetermined, however a probable cause can be attributed to prying/leveraging the anchor during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURE ANCHOR, MICRO CORKSCREW FT
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14074689
Report Number1220246-2022-04709
Device Sequence Number11964019
Product Code HWC
UDI-Device Identifier00888867008519
UDI-Public00888867008519
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K112237
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberSUTURE ANCHOR, MICRO CORKSCREW FT
Device Catalogue NumberAR-1318FT
Device Lot Number10358431
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/22/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-