• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIVATOR II; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIVATOR II; SNARE, FLEXIBLE Back to Search Results
Model Number M00561291
Medical Device Problem Codes Failure to Deliver Energy (1211); Failure to Cut (2587); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a captivator extra large round stiff snare was used during a colonoscopy procedure performed on (b)(6) 2022.It was reported that during the procedure and inside the patient, hot snaring was performed and it would not cut the tissue.The nurse applied too much pressure which caused the snare to bend and break near the handle.They tried another erbe diathermy machine but this did not resolve the problem even though the snare was securely attached to the active cord and no problems were noted with the cautery pins.The procedure was completed with a cold snaring technique with the use of a different captivator ii snare.There were no patient complications reported as a result of this event.Note: a photo was provided that showed the erbe device without any alert.Additionally, the photo showed the working length and wire was bent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIVATOR II
Common Device Name
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14074792
Report Number3005099803-2022-01818
Device Sequence Number14111686
Product Code FDI
UDI-Device Identifier08714729645801
UDI-Public08714729645801
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K131700
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/08/2024
Device Model NumberM00561291
Device Catalogue Number6129
Device Lot Number0026942101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/15/2022
Initial Report FDA Received Date04/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-