| Model Number |
DRV,3.5MM HEX,CANN (MINI HUDSON) |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported on (b)(6) 2022 by a facility representative via sems that an ar-8967d driver broke.This occurred during a case and broke during use.
|
| |
|
Search Alerts/Recalls
|