• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX SACETT SUCTION ABOVE CUFF ENDOTRACHEAL TUBE; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX SACETT SUCTION ABOVE CUFF ENDOTRACHEAL TUBE; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 100/189/070
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use of the product, leakage of air from the pilot balloon was observed.No patient injury was reported.
 
Additional Manufacturer Narrative
Other, other text: one device was returned for investigation.A pressure decay test found that there was a leak in the pilot balloon.Production personnel perform 100% visual inspection to inspect the pilot balloon and must be free from holes or any damage.Based on sample analysis, it was confirmed that there are no sharp surfaces that could cause the hole that is shaped as a pin hole which indicates that the damage was caused after product manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX SACETT SUCTION ABOVE CUFF ENDOTRACHEAL TUBE
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-15-26 fukushima,
minneapolis, MN 55442
MDR Report Key14075417
Report Number3012307300-2022-06173
Device Sequence Number2647432
Product Code BTR
UDI-Device Identifier15019315059520
UDI-Public15019315059520
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K081086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100/189/070
Device Catalogue Number100/189/070
Was Device Available for Evaluation? No
Date Returned to Manufacturer03/23/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2022
Supplement Date Received by Manufacturer05/16/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date06/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-