• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX SOFT SEAL POLAR MURPHY EYE TRACHEAL TUBE; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX SOFT SEAL POLAR MURPHY EYE TRACHEAL TUBE; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 100/136/050
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the pre-use check, the customer found difficulty inflating the cuff.No involvement was reported.
 
Additional Manufacturer Narrative
Other, other text: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Five (5) photos and a product sample were received for evaluation.Visual and functional testing was performed.Visual inspection found a cut on the tip of the tube was observed in photo 1 and inspection of the sample a cut on tip of the tube was detected.Functional testing found leakage was detected, the complaint was confirmed.The root cause of the reported issue was found to be the end form tube bevel operation not followed properly during manufacturing.Awareness notification was made to production personnel to explain the importance to adherence or following in the procedure end form tube bevel operation, by quality engineer on 25/apr/2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX SOFT SEAL POLAR MURPHY EYE TRACHEAL TUBE
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-15-26 fukushima,
minneapolis, MN 55442
MDR Report Key14075585
Report Number3012307300-2022-06175
Device Sequence Number17676404
Product Code BTR
UDI-Device Identifier15019315021053
UDI-Public15019315021053
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K931735
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2026
Device Model Number100/136/050
Device Catalogue Number100/136/050
Device Lot Number4080567
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/30/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2401-2021
Patient Sequence Number1
Patient SexUnknown
-
-