| Model Number |
1011-0139-05 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the associated defibrillator failed self test using these internal handle paddles.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned to zoll medical corporation.The customer's report was not replicated or confirmed.The internal handles were put through extensive testing using a simulator without duplicating the report.The internal handles were scrapped and replaced.Analysis of reports of this type has not identified an increase in trend.
|
| |
|