The following sections were updated in follow-up 1.B4, d9, g1, g3, g6, h2, h3, h6, h10.One 13.5f guidestar steerable guiding sheath was received back from the customer with the dilator.There were no other accessories.Blood was found on and inside the sheath and dilator.According to the event description summary, it took a lot of force to retract the balloon from the sheath and when and once released, a very loud pop sound was heard.Leakage from the valve was noticed toward the end of the procedure.Upon evaluation of the returned product, it was found that the sheath would not deflect when the deflection collar was rotated.The sheath was x-rayed through distal tip.It was found that the anchor ring remained in its original intended position and the pullwire was still attached without any damage.When the handle was x-rayed, it was found that the small and large hypotubes were intact.No slack was found in the pullwire that protrudes from the large hypotube when viewed under x-ray.When the handle was removed, it was found that the crimps that hold the hypotube/pullwire assembly together were insufficient and that's why the sheath no longer deflected.No breaks were found in the pullwire when viewed under x-ray.The valve was viewed under a 10x microscope and looked normal with no anomalies.The sheath was manually leak tested with the syringe connected to the stopcock and the tip occluded.The sheath did not exhibit leakage when tested using this method.The sheath was then tested using the iso 11070 for liquid leakage test method.The device was occluded at the distal tip and pressurized to 38kpa and held at this pressure for 30 seconds and no leakage was observed from the hemostatic valve.The sheath was further pressurized beyond 300kpa and no leakage was observed from the stopcock, sideport tubing or through the handle.A fit check was performed with the accompanying 13.5f dilator.The dilator passed fully and smoothly all the way through the sheath.The device history records were reviewed to confirm that the device passed all applicable in-process and final inspections.Deflection issue: returned device analysis revealed sheath would no longer deflect.The crimp in the hypotube that holds the pullwire in place was not sufficient enough to hold the entire assembly together.The reason for return was confirmed.The crimping of the hypo-tube portion of the handle assembly process for this packaged lot is from 4 different handle assembly lots.Process deviation was initiated to ensure the crimp on the hypotube assembly holds together.Manufacturing and qa personnel have been trained to the changes stated in the deviation.Ccr210813-a was completed to update the procedure to implement a dual crimper.Capa was initiated to investigate the root cause and implement corrective action to prevent recurrence of this issue.Note: the force it took in order to withdraw the balloon catheter along insufficient crimping are most likely the reason the sheath would no longer deflect.It is unknown why the user heard a popping sound as the pullwire did not break.Leakage issue: returned device analysis revealed the sheath was within manufacturing specifications.The hemostatic valve looked normal and intact.The sheath did not leak when tested manually.The sheath also met the iso leakage test method requirement.A fit check using the accompanying 13.5f dilator confirmed that the inner diameter of the sheath was within spec.According to the dhr, the sheath passed all in-process and final inspection steps including visual, mechanical, dimensional and leak testing.No manufacturing defects were found.Per manufacturing procedure (destino twist steerable guiding sheath handle assembly) sample size: 100%.At the time of manufacture, the procedure used to assemble this unit was at rev-02.The following procedure was used to assemble.Position the small hypo-tube onto the pull wire of the sheath.Insert the small hypo-tube into the large hypo tube until it is bottomed out inside the large hypo-tube.Using a ruler, measure the exposed length of the small hypo-tube.Pull the small hypotube an additional 4+/-0.5mm out of the large hypo-tube.Verify that the exposed length of the small hypo-tube is 4+/-0.5mm longer than the bottomed out measurement.Verify that there are two crimps present and using a ruler verify that the distance between the crimp 1 and the proximal end is 3+/-0.5mm and the second crimp is 8+/-0.5mm from the proximal end.Trim the excess pull wire that extends past the small hypo-tube.The procedure was updated as part of ccr210813-a to add the following controls: to add dual crimper method for 12f and larger sheaths.Manually deflect the sheath by grabbing the shaft with one hand about 4 inches from the distal tip and grabbing the worm gear with the other hand and pull back on the worm gear.In the event the crimp lets loose during deflection, rework the crimping process.Per qa procedure: (destino steerable guiding sheath in-process and final inspection) perform deflection test according to procedure.The deflection test is performed by manufacturing personnel at 100% and inspected by quality assurance personnel.Aql sample size: ansi z 1.4, gen level i, normal, aql 1.5 normal leak test: sample size: ansi z 1.4, gen level i, normal, aql 1.5 normal perform leak test according to procedures based on available leak tester.The leak test is performed by manufacturing personnel at 100% and is observed by quality assurance personnel at aql level ansi z 1.4, gen level i, normal, aql 1.5 normal.Per ifu: preparing steerable sheath for insertion verify deflecting and straightening of the distal section of the steerable sheath using the handle of the sheath.Refer to deflecting and straightening the steerable sheath" for instructions.General use of the steerable sheath do not force the steerable sheath assembly if significant resistance is encountered during the insertion or passage.If resistance is encountered, determine the cause and correct before continuing the procedure.The steerable sheath must be thoroughly flushed with either saline or heparinized saline and free of air prior to use to avoid air embolism to the patient.Aspiration and flushing of the sheath should be performed frequently to help minimize the potential for air embolism.For injecting or aspirating through the sheath, use the sideport only with stopcock.Prior to infusion, remove all air using the sideport.No further follow-up is required.Based on the investigation, a capa has been initiated to investigate and prevent the reoccurrence of this issue in guidestar sheath.In addition, there was no new failure mode identified and the risk remains acceptable.The event will be re-evaluated if additional information becomes available.Oscor will continue to monitor this event type and risk.Oscor inc.Is submitting the above report to comply with 21 cfr part 803, the medical device reporting regulation.The report is based upon information obtained by oscor inc.Which the company may not have been able to investigate or verify prior to the date the report was required by fda.This report does not constitute an admission that the device, oscor inc., or its employees, caused or contributed to the event described in this report.Nor does this report reflect a conclusion by fda, oscor inc., or its employees, that the report constitutes an admission that the device, oscor inc., or its employees caused or contributed to the event described in this report.
|