• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 SCREW 6.5MM X 20MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 SCREW 6.5MM X 20MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: foreign country: (b)(6).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2021-02498.Visual evaluation of the returned product found the seals do not meet the minimum requirement and are non- conforming.Sterility has been compromised.Review of the device history record(s) identified no deviations or anomalies during manufacturing.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported by the distributorship that the products were found to be nonconforming.The incoming inspection team member found the seal width was less than the accepted tolerance.No adverse events have been reported as a result of the malfunction.No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 SCREW 6.5MM X 20MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14078642
Report Number0001825034-2022-00871
Device Sequence Number2122393
Product Code PBI
UDI-Device Identifier00880304527454
UDI-Public(01)00880304527454(17)310525(10)7036082
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number010000997
Device Lot Number7036082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2021
Initial Report FDA Received Date04/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-