| Model Number |
N/A |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Report source: foreign country: (b)(6).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2021-02498.Visual evaluation of the returned product found the seals do not meet the minimum requirement and are non- conforming.Sterility has been compromised.Review of the device history record(s) identified no deviations or anomalies during manufacturing.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported by the distributorship that the products were found to be nonconforming.The incoming inspection team member found the seal width was less than the accepted tolerance.No adverse events have been reported as a result of the malfunction.No further event information available at the time of this report.
|
| |
|