• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL GALAXY G3 MINI 2MM X 4CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL GALAXY G3 MINI 2MM X 4CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number GLM920040
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 03/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient date of birth was not provided.Procode is krd/hcg.[conclusion]: the event was reported via the sterling study, a 71-year-old male patient with a history of class 2 obesity (bmi 35 kg/m2) underwent coil embolization of a ruptured left anterior communicating artery (acom) aneurysm with hunt & hess grade 1 on (b)(6) 2021.The dimensions of the aneurysm were as follows: aneurysm height 5.9mm, dome 3.5mm, maximum aneurysm diameter 5mm, neck size 2.4mm, and dome-to-neck ratio of 1.5mm.The parent vessel diameter was 1.9mm.Coil embolization was performed with six cerenovus spectra coils: a 5 mm x 15 cm galaxy g3 (gly120515 / 30513182), a 3 mm x 8 cm galaxy g3 xsft (glx120308 / l14664), a 3 mm x 6 cm galaxy g3 xsft (glx120306 / 30490139), a 2 mm x 6 cm galaxy g3 mini (glm920060 / 30392250), a 1.5 mm x 2 cm galaxy g3 mini (glm915020 / l14683), and a 2 mm x 4 cm (glm920040 / 30379814) via an sl-10® microcatheter (stryker).Microcatheter kickback (loss of access to aneurysm) was experienced three times during the procedure.Heparin was administered intraoperatively.In the opinion of the investigator, treatment of the target aneurysm was considered complete, and the study coils were successfully implanted at the target site with 69% angiosuite packing density.Immediate post-procedure modified raymond-roy classification score was class ii: residual neck = persistence of any portion of the original defect of the arterial wall but without opacification of the aneurysmal sac.There were no reported study device deficiencies or intraoperative complications.The patient was discharged home with self-care on (b)(6) 2021.On (b)(6) 2022, he underwent aneurysm retreatment due to residual neck.Coils (unspecified) and a 3mm x 21mm atlas stent (stryker) were implanted with final modified raymond-roy classification score of class i: complete obliteration.On the following day, the patient developed transient speech impairment.Head computed tomography (ct) scan was obtained and no acute findings were noted.Symptoms resolved later that day.Per the principal investigator (pi), the event was moderate in severity and not related to the study device.The event was not considered serious.The patient was discharged to a rehabilitation center on (b)(6) 2022.The complaint coils remain implanted in the patient and are thus not available for evaluation.On (b)(6) 2022, additional information was provided.The information indicated that the aneurysm was found to have filling of the neck and body at follow-up angiography.Recanalization was diagnosed at 60% aneurysm occlusion.The coils remained in the aneurysm sac.There was evidence of coil compaction.The patient did not experience any clinical symptoms as a result of the aneurysm recanalization.Based on complaint information, the device remains implanted and is thus not available for evaluation.A review of manufacturing documentation associated with this lot (30379814) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.The device will not be returned for analysis and therefore, no further investigation can be performed at this time.As a result, we are closing this investigation.Coil compaction is the decrease in interspaces between the loops of the coils, which leads to smaller coil mesh.It occurs over time after coil placement.Return of blood flow into the aneurysm due to coil compaction may increase the risk for aneurysm rupture and possibly require additional intervention.The reported coil compaction and aneurysm recanalization necessitated endovascular retreatment to preclude permanent damage or injury.Furthermore, the relationship of the study coils to the reported event cannot be excluded (i.E., coil compaction).Thus, the event is considered serious and mdr reportable for all implanted cerenovus coils.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This is one of 6 products involved with the reported complaint.The associated manufacturer report numbers are: 3008114965-2022-00240, 3008114965-2022-00241, 3008114965-2022-00242, 3008114965-2022-00243, 3008114965-2022-00244.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The event was reported via the sterling study, a 71-year-old male patient with a history of class 2 obesity (bmi 35 kg/m2) underwent coil embolization of a ruptured left anterior communicating artery (acom) aneurysm with hunt & hess grade 1 on (b)(6) 2021.The dimensions of the aneurysm were as follows: aneurysm height 5.9mm, dome 3.5mm, maximum aneurysm diameter 5mm, neck size 2.4mm, and dome-to-neck ratio of 1.5mm.The parent vessel diameter was 1.9mm.Coil embolization was performed with six cerenovus spectra coils: a 5 mm x 15 cm galaxy g3 (gly120515 / 30513182), a 3 mm x 8 cm galaxy g3 xsft (glx120308 / l14664), a 3 mm x 6 cm galaxy g3 xsft (glx120306 / 30490139), a 2 mm x 6 cm galaxy g3 mini (glm920060 / 30392250), a 1.5 mm x 2 cm galaxy g3 mini (glm915020 / l14683), and a 2 mm x 4 cm (glm920040 / 30379814) via an sl-10® microcatheter (stryker).Microcatheter kickback (loss of access to aneurysm) was experienced three times during the procedure.Heparin was administered intraoperatively.In the opinion of the investigator, treatment of the target aneurysm was considered complete, and the study coils were successfully implanted at the target site with 69% angiosuite packing density.Immediate post-procedure modified raymond-roy classification score was class ii: residual neck = persistence of any portion of the original defect of the arterial wall but without opacification of the aneurysmal sac.There were no reported study device deficiencies or intraoperative complications.The patient was discharged home with self-care on (b)(6)2021.On (b)(6) 2022, he underwent aneurysm retreatment due to residual neck.Coils (unspecified) and a 3mm x 21mm atlas stent (stryker) were implanted with final modified raymond-roy classification score of class i: complete obliteration.On the following day, the patient developed transient speech impairment.Head computed tomography (ct) scan was obtained and no acute findings were noted.Symptoms resolved later that day.Per the principal investigator (pi), the event was moderate in severity and not related to the study device.The event was not considered serious.The patient was discharged to a rehabilitation center on (b)(6) 2022.The complaint coils remain implanted in the patient and are thus not available for evaluation.On (b)(6) 2022, additional information was provided.The information indicated that the aneurysm was found to have filling of the neck and body at follow-up angiography.Recanalization was diagnosed at 60% aneurysm occlusion.The coils remained in the aneurysm sac.There was evidence of coil compaction.The patient did not experience any clinical symptoms as a result of the aneurysm recanalization.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY G3 MINI 2MM X 4CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CODMAN & SHURTLEFF, INC. (FREMONT)
47709 fremont blvd
fremont CA 94538
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key14078937
Report Number3008114965-2022-00245
Device Sequence Number2096493
Product Code KRD
UDI-Device Identifier10886704080305
UDI-Public10886704080305
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K171862
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberGLM920040
Device Catalogue NumberGLM920040
Device Lot Number30379814
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2022
Initial Report FDA Received Date04/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/06/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GALAXY G3 5MM X 15CM; GALAXY G3 MINI 1.5MM X 2CM; GALAXY G3 MINI 2MM X 6CM; GALAXY G3 XSFT 3MM X 6CM; GALAXY G3 XSFT 3MM X 8CM; SL-10® MICROCATHETER (STRYKER); SYNCHRO2® 0.014¿ GUIDEWIRE (STRYKER)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
Patient Weight104 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-