| Model Number |
640CX2505 |
| Medical Device Problem Code |
Stretched (1601)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Procode is krd/hcg.(b)(6).The initial reporter email was not reported / available.The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.The photos included in the complaint underwent review from the product analysis lab.[photo analysis]: the two photos showing the complaint device were attached to the complaint file.The reported condition regarding the embolic coil being stretched was confirmed based on the photos.The middle part of the embolic coil component was noted to be in a severely stretched condition; it is still attached to the rest of the delivery unit since the stretch-resistant fiber is in one piece.This damage might have occurred when the embolic coil was retracted, however, no further evaluation can be performed at this time.A review of manufacturing documentation associated with this lot (30615576) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.The manufacturing record evaluation (mre) was performed.The reported failure is not related to the manufacturing process.Further investigation will be performed when the device is returned for evaluation and analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
The healthcare professional reported that during an endovascular coil embolization procedure, the microcatheter (unspecified brand) was placed in the target position, the physician filled the aneurysm with the coil.The physician was not satisfied with the shape of the coil and tried to retract the coil, but it was found to be stretched inside the microcatheter.A new device was used as replacement to complete the procedure.There was no report of any patient injury or adverse event.The complaint included photos of the complaint device.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the product was received the product analysis lab on 27-jun-2022.The return product is pending evaluation.A supplemental 3500a report will be submitted once the product investigation has been completed.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer¿s ref.(b)(4).The purpose of this mdr submission is to report the investigational finding of the returned device.[conclusion]: the healthcare professional reported that during an endovascular coil embolization procedure, the microcatheter (unspecified brand) was placed in the target position, the physician filled the aneurysm with the 2.5mm x 5cm orbit galaxy complex xtrasoft coil (640cx2505 / 30615576).The physician was not satisfied with the shape of the coil and tried to retract the coil, but it was found to be stretched inside the microcatheter.A new device was used as replacement to complete the procedure.There was no report of any patient injury or adverse event.The complaint included photos of the complaint device.Multiple attempts to obtain additional information related to the procedure and the reported device issue were unsuccessful.If additional information is received at a later date, this file will be updated accordingly.The photos included in the complaint underwent review from the product analysis lab.One of the photos that accompanied the complaint file shows the coil stretched and partially outside from the introducer.No other damages can be observed.The complaint device was returned for evaluation and analysis.The investigation finding is documented below.Investigation summary: a non-sterile 2.5mm x 5cm orbit galaxy complex xtrasoft coil was received contained the decontamination pouch.Visual inspection was performed.It was noted that the embolic coil was partially outside of the introducer.No other damages nor anomalies were observed on the returned complaint device during the visual inspection.Microscopic inspection was performed.Under magnification, the embolic coil component was noted to be severely stretched while still remaining attached to the rest of the device.The stretched condition observed on the embolic coil appeared to be more severe than the condition shown in one of the two photos included in the complaint.This damage may have occurred during post-operative handling of the device, during shipping, or during the decontamination process.The reported issue documented in the complaint was confirmed based on the microscopic inspection of the returned device.The embolic coil was noted to be in a severely stretched condition.However, based on the limited information available, the exact time when this condition occurred cannot be conclusively determined.Coil stretching is a known potential issue associated with the use of this device.The instructions for use (ifu) provides proper handling instructions for the device to prevent such issues from occurring.Stretching can occur during procedure handling where force may have been inadvertently applied.It is possible that during the retraction of the coil, undue force was inadvertently applied, or that an adequate flush had not been maintained through the microcatheter.However, based on the limited information available, the root cause remains unknown and speculative.A review of manufacturing documentation associated with this lot (30615576) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|
| |
|