| Model Number |
720054-02 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision surgery in which a spectra concealable penile prosthesis (spp) was removed and replaced with a tactra due to incorrect device size.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the returned device did not confirm the as reported incorrect size.The cylinders were visually inspected, microscope examined and functionally tested.However, analysis did identify that both cylinders had a sharp instrument/tool damage to the outer tube in the proximal cylinder bodies, which resulted in a hole and exposed metal segments.This damage is consistent with explant and will be considered a secondary failure.The cylinders performed within specifications.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision surgery in which a spectra concealable penile prosthesis (spp) was removed and replaced with a tactra due to incorrect device size.No further information was provided.
|
| |
|