• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM PASSIVE FIXATION, DF-4 CONNECTOR; DEFIBRILLATION LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM PASSIVE FIXATION, DF-4 CONNECTOR; DEFIBRILLATION LEAD Back to Search Results
Model Number 7171Q/65
Medical Device Problem Codes Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291); Over-Sensing (1438); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the asymptomatic patient presented for a follow up in clinic with a right ventricular lead that exhibited lead noise, r-wave amplitude variation, and high pacing impedance due to suspected conductor fracture.Lead revision was discussed; no interventions were reported.The patient was in stable condition.
 
Additional Manufacturer Narrative
The reported events of fracture, sensing noise, r-wave amplitude variation, and high pacing impedance were confirmed.As received, a partial lead was returned in two pieces.Visual and x-ray examination found a possible inner coil fracture at the middle region of the lead.Additional analysis found that all four filars of the inner coil were fractured.The cause of the reported events was due to fracture of inner coil consistent with fatigue fracture.
 
Event or Problem Description
New information received noted that the lead noise was over-sensed.The lead was explanted and replaced.The patient was in stable condition post-procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM PASSIVE FIXATION, DF-4 CONNECTOR
Common Device Name
DEFIBRILLATION LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14079310
Report Number2017865-2022-07660
Device Sequence Number17461967
Product Code NVY
UDI-Device Identifier05414734503297
UDI-Public05414734503297
Combination Product (Y/N)Y
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2010
Device Model Number7171Q/65
Device Catalogue Number7171Q-65
Device Lot Number2863696
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2022
Supplement Date Received by Manufacturer04/13/2022
06/28/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date04/14/2022
06/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/07/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ELLIPSE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexMale
Patient Weight91 KG
-
-