BIOTRONIK SE & CO. KG SENTUS PROMRI OTW QP L-85; LV LEAD Back to Search Results
Model Number 401183
Medical Device Problem Codes Fracture (1260); High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that pacing impedance on this lv lead was out of range (greater than 3000 ohm) during the follow-up approx.11 months after the implantation.Lead breakage was suspected.The lead was explanted and replaced.
 
Additional Manufacturer Narrative
Upon receipt, the lead under complaint was subjected to an extensive analysis.The performance of the lead was scrutinized, including a visual, mechanical and electrical inspection.The analysis showed that the conductor coil was found fractured 48 cm distal to the is4 connector pin, which is assumed to be the root cause of the reported clinical observations.Based on the characteristics, as well as the location of the fracture, it is reasonable to assume that the lead was subject to excessive mechanical forces as the result of clavicular-first rib entrapment.Furthermore, cuts into the insulation were found 38 cm distal to the is4 connector pin, which most likely resulted from the extraction procedure.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTUS PROMRI OTW QP L-85
Common Device Name
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key14079443
Report Number1028232-2022-01849
Device Sequence Number11293065
Product Code OJX
UDI-Device Identifier04035479138448
UDI-Public04035479138448
Combination Product (Y/N)N
PMA/510(K) Number
P070008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model Number401183
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2022
Supplement Date Received by Manufacturer06/02/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-