| Model Number |
M00536520 |
| Medical Device Problem Codes |
Difficult to Advance (2920); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an advanix pancreatic stent was used during an endoscopic pancreatic duct stent implantation procedure in the pancreatic duct, performed on (b)(6) 2022.During the procedure, when the physician attempted to deploy the stent, severe resistance was felt between the guide catheter and the stent.The stent shrank and was deformed upon deployment.The stent was successfully implanted to the patient and completed the procedure.A photo of the complaint device was provided and showed that the stent was deformed.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The complainant indicated that the device remains implanted and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|