| Model Number |
AVEA |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to vyaire medical that the avea ventilator is auto cycling.The customer confirmed that there was no patient involvement associated with the reported event.The issue occurred during set-up, prior to patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: g3, g6, h2, h3, h6 and h10 result of investigation: vyaire field service went onsite confirmed unit is auto cycling.Peep (positive end expiratory pressure) drop to 0 and auto cycle stops.No damage in exhalation corner.As a resolution reseated gde (gas delivery engine) ran inspiratory and expiratory calibration.Ran expiratory flow calibration all calibration passed.Ran est (extended system test) passed.Unit still auto cycling.
|
| |
|