• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS SPECTRA CONCEALABLE PENILE PROSTHESIS; PROSTHESIS PENILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS SPECTRA CONCEALABLE PENILE PROSTHESIS; PROSTHESIS PENILE Back to Search Results
Model Number 720054-01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 02/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this spectra penile prosthesis (spp) experienced pain at the surgical site, which led to a surgical intervention.During the procedure, the spp was removed and replaced with a tactra malleable penile prosthesis.There were no additional patient complications reported.
 
Additional Manufacturer Narrative
Device analysis pending closure.
 
Additional Manufacturer Narrative
Upon receipt at our quality assurance laboratory, this spp underwent a thorough analysis.The cylinders were visually inspected, microscopically examined, and functionally tested.Both cylinders had sharp instrument damage which likely occurred during explant; besides that, no anomalies were found.The cylinders performed within specification.Product analysis was unable to confirm the clinical observations.A review of the labeling confirmed the reported adverse events of pain is included in the product labeling.Additionally, there is no objective evidence that the user did not properly handle or use the device according to the spectra concealable penile prosthesis instructions for use (ifu).The reported events also do not contain an allegation against the labeling.
 
Event or Problem Description
It was reported that the patient with this spectra penile prosthesis (spp) experienced pain at the surgical site, which led to a surgical intervention.During the procedure, the spp was removed and replaced with a tactra malleable penile prosthesis.There were no additional patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS SPECTRA CONCEALABLE PENILE PROSTHESIS
Common Device Name
PROSTHESIS PENILE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
4089353452
MDR Report Key14079803
Report Number2124215-2022-10725
Device Sequence Number17528598
Product Code FAE
UDI-Device Identifier00878953005201
UDI-Public00878953005201
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K090663
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/25/2025
Device Model Number720054-01
Device Catalogue Number720054-01
Device Lot Number1000428843
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer05/19/2022
Initial Report FDA Received Date04/11/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexMale
-
-