Manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned for evaluation.Based on the investigation performed in the laboratory, the stent graft was stuck inside the sheath and could not be further deployed.Therefore, a partial deployment could be confirmed.The system was flushed, and appropriate introducer sheath and guidewire were used.The definitive root cause could not be determined based upon available information.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.Regarding the preparation of the device the instructions for use states carefully remove the endovascular system from its packaging and inspect packaging and system for any damage or defects.'flush the stent graft lumen with sterile saline by using a small volume syringe.A) attach the syringe to the luer port at the back of the endovascular system (.).B) attach the syringe to the luer port on the y-adapter, (.) close the stopcock when flushing is complete and remove the syringe from the luer port'.Regarding the precautions prior and during deployment the instructions for use states 'prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure', ¿higher deployment force may be encountered (¿) in tortuous anatomy¿, and 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.¿.The instructions for use states regarding the accessories 'the use of an appropriately sized introducer sheath is recommended; prepare a stiff 0.035" guidewire per its instructions for use and advance the guidewire under fluoroscopy to the target location'.The packaging pictograms indicate an introducer size of 9f and a 0.035" guidewire.(expiry date: 07/2024).
|