| Model Number |
300-30-06 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Bone Fracture(s) (1870)
|
| Date of Event |
03/23/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, during stem impaction in a l total shoulder procedure on a (b)(6) female patient, there was a calcar fracture to the l shoulder.It was repaired with a cerclage and the procedure was completed.The patient has a history of osteoarthritis.There is no device to be returned.The case report form indicates this event is possibly related to devices and/or possibly related to procedure.This event report was received through clinical data collection activities.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending evaluation.Concomitant device(s): 310-01-41 - equinoxe, humeral head short, 41mm (alpha), 300-10-45 - equinoxe replicator plate 4.5mm o/s, and 314-13-02 - equinoxe cage glenoid small, alpha.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) the calcar fracture was reported to have occurred during the implant implementation.It likely occurred while preparing the shoulder for and may have been related to the feed rate/pressure of the calcar planer, reamer rotational speed, or the patient¿s bone quality/geometry.
|
| |
|