• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number RONYX30022UX
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Insufficient Information (4580)
Event Date 04/08/2022
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
During a procedure an attempt was made to use one resolute onyx rx coronary drug eluting stent to treat a non-tortuous, severely calcified lesion exhibiting 90% stenosis located in the proximal left main (lm) coronary artery.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was encountered when advancing the device.It was reported that stent dislodgement occurred during removal following a failed delivery.The dislodged stent was not removed from the patient and was crushed into vessel wall.The patient was reported to be alive with no further injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14083047
MDR Text Key289060116
Report Number9612164-2022-01411
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier00643169557000
UDI-Public00643169557000
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRONYX30022UX
Device Catalogue NumberRONYX30022UX
Device Lot Number0011072067
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/08/2022
Initial Date FDA Received04/12/2022
Supplement Dates Manufacturer Received06/01/2022
Supplement Dates FDA Received06/07/2022
Date Device Manufactured02/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age78 YR
Patient SexMale
-
-