• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH LARGE EX-FIX OPEN ADJ CLAMP MR-CONDITIONA; APPLIANCE, FIXATION, NAIL/BLADE/PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH LARGE EX-FIX OPEN ADJ CLAMP MR-CONDITIONA; APPLIANCE, FIXATION, NAIL/BLADE/PLATE Back to Search Results
Catalog Number 390.008
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was determined on an unknown date five ex-fix clamps were unable to connect.It was stated the rod does not fit well.Patient and surgical involvement is unknown.No serious injury reported.This is report 3 of 5 for (b)(4).This report is for a large ex-fix clamp.
 
Additional Manufacturer Narrative
Additional narrative: additional code: ktt.Complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter is j&j company representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h4, h6: a device history record (dhr) review was conducted: part #: 390.008 lot #: 6185647 manufacturing site: werk selzach supplier: synthes usa hq, inc release to warehouse date: 29 sep 2009 expiration date: n/a a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE EX-FIX OPEN ADJ CLAMP MR-CONDITIONA
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14083389
Report Number8030965-2022-02352
Device Sequence Number5317735
Product Code LXT
UDI-Device Identifier07611819760592
UDI-Public(01)07611819760592
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K122455
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number390.008
Device Lot Number6185647
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/14/2022
Supplement Date Received by Manufacturer04/21/2022
Initial Report FDA Received Date04/12/2022
Supplement Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/29/2009
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAMP CAN BE MOUNTED SELF-HOLD; CLAMP CAN BE MOUNTED SELF-HOLD; CLAMP CAN BE MOUNTED SELF-HOLD; CLAMP CAN BE MOUNTED SELF-HOLD; UNKNOWN RODS
Patient SexUnknown
-
-