| Model Number |
27040XB |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The ceramic tip fractured/broke from the sheath at end of procedure when pulling out instrument.A missing piece is still in patient.The sheath is over 8 years old.
|
| |
|
Event or Problem Description
|
|
During a turp, surgeon completed the surgery, and realized when pulling out the instruments that a broken piece of ceramic came out.A missing piece is still in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation report from kst/mfr.Was received which states: the conclusion and root cause of the non-return product: "the most likely cause is the age ( 8 years) of the item and the high" application cycles.This probably caused micro cracks on the ceramic beak, which led to the breakage of the beak during extraction.
|
| |
|