• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number RLT261418
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 03/16/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: gore® excluder® conformable aaa endoprosthesis with active control system (cxa280005e ¿ lot: 23550552).Device remains implanted.The initial imaging evaluation stated the following: there is no contrast visualized outside the implanted endograft for the proximal 25mm.The graft appears to be well apposed to the aorta within this location.Blood is visualized outside the endograft >25mm distal to the proximal portion of the main body.This appears to be a well-timed cta examination in the arterial phase.There is no contrast visualized within the venous structures.The jpeg images provided are non-dicom, reviewed and not added to this evaluation as they add no value and are non-dicom with no patient information or scan details.Additional imaging of patient after reintervention to be sent for investigation when available.Statement from ifu: "adverse events that may occur and / or require intervention or additional intraoperative procedure time include, but are not limited to endoleak." review of imaging shows graft appeared to be well apposed to the aorta within this location.A review of the manufacturing records indicated the lot met all pre-release manufacturing specifications.
 
Event or Problem Description
It was reported that a patient presented with an abdominal aortic aneurysm and was treated with a gore® excluder® aaa endoprosthesis on (b)(6) 2017.The patient presented with a type 1a endoleak on march 16, 2022.The patient was treated for endoleak with a gore® excluder® conformable aaa endoprosthesis with active control system (cxa280005e ¿ lot: 23550552) on (b)(6) 2022.After two dilatations, no type i endoleak was visible angiographically.The patient tolerated the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
tom hormby
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key14084586
Report Number3007284313-2022-01861
Device Sequence Number12059820
Product Code MIH
UDI-Device Identifier00733132622474
UDI-Public00733132622474
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/20/2020
Device Model NumberRLT261418
Device Catalogue NumberRLT261418
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer09/29/2022
Initial Report FDA Received Date04/12/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/21/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Other;
Patient Age78 YR
Patient SexMale
-
-