ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE FIXED BEARING POSTERIOR STABILIZED (PS) LEFT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Codes Appropriate Device Problem Term/Code Not Available (3191); Noise, Audible (3273)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060)
Date of Event 03/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a revision approximately 3 years post implantation due to pain with activities, clicking, and limited extension.During the revision, significant scarring and heterotrophic bone formation was debrided.A new articular surface was placed without complication.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: 42532007101 - tibial component - 64110531, 42540200032 - patella - 64291052, 42500606401 - femoral component - 63946879, 42557000114 - stem extension - 64358761.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2022-00092 and 0002648920-2022-00086.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified that the patient underwent a revision for pain and limited rom.Extensive scar tissue and heterotopic bone formation were seen on the tibial and patella bone.No infection was noted.No loosening was noted in the x-rays.A definitive root cause cannot be determined.Per package insert, pain and poor range of motion are known adverse effects of the system.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE FIXED BEARING POSTERIOR STABILIZED (PS) LEFT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14084766
Report Number3007963827-2022-00091
Device Sequence Number2134353
Product Code MBH
UDI-Device Identifier00889024237247
UDI-Public(01)00889024237247(17)220531(10)63689140
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K121771
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model NumberN/A
Device Catalogue Number42512400710
Device Lot Number63689140
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer06/22/2022
Initial Report FDA Received Date04/12/2022
Supplement Report FDA Received Date06/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/22/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight99 KG
-
-