| Model Number |
826851 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
While demonstrating how to use the cerelink monitor an error message appeared saying that the system detected a problem either with the extension cable or sensor.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The cerelink microsensor was returned for evaluation: dhr: lot: 3362996 (sn: (b)(6) met specifications when released.Failure analysis - no visible damage to the millar sensor, catheter or connector.Cerelink monitor reading was acceptable; icp express passed with reading 511.The device passed electronic, noise, linearity/hysteresis, and signal drift tests.The issue of the complaint was not confirmed.The root cause of the issue reported by customer could not be determined as the device worked correctly.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|