• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD UNKNOWN CERELINK; ICP MICRSOSENSORS - CERELINK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD UNKNOWN CERELINK; ICP MICRSOSENSORS - CERELINK Back to Search Results
Model Number 826851
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
While demonstrating how to use the cerelink monitor an error message appeared saying that the system detected a problem either with the extension cable or sensor.No patient involvement.
 
Additional Manufacturer Narrative
The cerelink microsensor was returned for evaluation: dhr: lot: 3362996 (sn: (b)(6) met specifications when released.Failure analysis - no visible damage to the millar sensor, catheter or connector.Cerelink monitor reading was acceptable; icp express passed with reading 511.The device passed electronic, noise, linearity/hysteresis, and signal drift tests.The issue of the complaint was not confirmed.The root cause of the issue reported by customer could not be determined as the device worked correctly.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN CERELINK
Common Device Name
ICP MICRSOSENSORS - CERELINK
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14085477
Report Number3013886523-2022-00152
Device Sequence Number2096693
Product Code GWM
UDI-Device Identifier10886704075431
UDI-Public10886704075431
Combination Product (Y/N)N
PMA/510(K) Number
NI
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Model Number826851
Device Catalogue NumberXXX - CERELINK
Device Lot Number101072
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer05/17/2022
Initial Report FDA Received Date04/12/2022
Supplement Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/08/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-