Model Number HWD 1000
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation of battery charger/modem has been completed.The reported problem (will not power on) has been confirmed.Upon investigation the battery charger/modem was unable to recharge a battery.The cause for the failure was isolated to a shorted q1 current-controlling transistor on the bedside pca.The root cause for the shorted q1 transistor could not be positively identified.No adverse event resulted from the damaged charger.
 
Event or Problem Description
A us distributor reported that a battery charger was unable to charge a battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HWD
Common Device Name
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
landyn blinn
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key14085832
Report Number3008642652-2022-03793
Device Sequence Number12084398
Product Code MVK
UDI-Device Identifier00855778005173
UDI-Public00855778005173
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P010030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberHWD 1000
Device Catalogue Number10A0997
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Initial Report FDA Received Date04/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1