Model Number 510-00-010
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994)
Date of Event 03/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - patient had pain and function problems.Surgery required due to stem loosening and fractures observed via x-ray.Glenoid appears fine.After assessment by surgeon osteolysis and fractures as a result, possible infection too.
 
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Additional Manufacturer Narrative
Corrected data: see a.2., b.2., d.9.& h.3.Manufacturer narrative: the reason for this revision surgery was reported as osteolysis, fractures and possible infection.The previous surgery and the surgery detailed in this event occurred 4.1 years apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were not released by hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to osteolysis, fractures and possible infection.There were no findings during this evaluation that indicate the reported device was the source of the osteolysis and fractures or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have caused the osteolysis, fracture or contributed to the possible infection or inhibited the patient's immune system.There are multiple factors that may also contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Common Device Name
RSP, PRIMARY HUMERAL MONOBLOCK, 10MM X 108
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, TX 78758-5445
MDR Report Key14085942
Report Number1644408-2022-00473
Device Sequence Number11299587
Product Code PHX
UDI-Device Identifier00888912024822
UDI-Public(01)00888912024822
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K100741
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/27/2023
Device Model Number510-00-010
Device Catalogue Number510-00-010
Device Lot Number980F1368
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer05/10/2022
Initial Report FDA Received Date04/12/2022
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/04/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-00-432 LOT 918F1961
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient Weight69 KG
-
-