| Model Number |
86623 |
| Medical Device Problem Code |
Operating System Becomes Nonfunctional (2996)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been received for analysis.It will not be returned to boston scientific for laboratory analysis since the issue has been resolved and it is currently working properly.If there is any further relevant information obtained, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure the labsystem pro pc crashed/froze and no electrocardiogram (ecg) signals visualization was possible.The software was closed and restarted to resolve the issue.A philips vital monitor was available for spo2 and hearth rate vitals.The procedure was successfully completed without any patient complications.The device is currently working and is not expected to be returned for analysis.
|
| |
|