Per clinical trial (b)(6): on (b)(6) 2018, patient underwent bilateral augmentation mastopexy using a full vertical and horizontal t incision.One piece of galaflex was used and cut into a rectangle.No concomitant procedures were performed.On (b)(6) 2019, left migration, malposition or contraction was reported.Note states that physician can do left lateral capsulorraphy to make breasts more even.No documentation of further procedure being done.No further information given and resolution date not given.Review by medical affairs identifies as likely related to the device and serious.
|
|
Based on the information provided, no conclusions can be made.As reported, patient experienced post-op complication of migration.Medical affairs has assessed the patient¿s postoperative course as being serious and likely related to the study device.Review of the device history records confirmed the device was manufactured to specification.Migration is a known inherent risk of the surgery/use of the device and listed as a possible complication in the adverse reaction section of the instructions for use (ifu) supplied with the device.Note: this mdr report is related to a retrospective study conducted to proactively gather real-world evidence on the use and safety of the p4hb product family, including galaflex scaffold, galaflex 3d, and galaflex 3dr.All safety events from the study are adjudicated by the bd medical monitor for an independent assessment of severity and relatedness.In an abundance of caution, events for which relatedness cannot be ruled out are processed for mdr reporting.Not returned - remains implanted.
|