Product complaint # (b)(4).Additional information was requested, but unavailable: what tissue was being approximated/ sutured? no further information is available.Lot number? no further information is available.Procedure name? no further information is available.No further information will be provided.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Product complaint # (b)(4).Date sent to the fda: 5/23/2022.Additional information: h3 investigational summary: visual analysis of the returned product revealed that one empty labeled winding former and a needle-suture piece product code z510g were received to ethicon for evaluation.Upon visual inspection of the returned sample, the suture was received with body fluids, fraying and the end was observed cut.After further evaluation crimping marks were observed on the surface suture possibly from a surgical instrument.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.The manufacturing records couldn't be reviewed as the batch number is unknown.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|